User Manual Druck DPI612
Page 49 of 105 [EN] English – 109M4017 Revision -
2.3.4 Priming Sequence
1 2 3 4
5 6 7 8
Step Procedure
Turn the Volume Adjuster counter-clockwise. (See Section
1.9.7)
Turn the Volume Adjuster 10 turns clockwise
3 Close Release Valve.
Close Refill Valve clockwise, finger tight (See Section 1.9.8).
Operate Priming Pump until liquid is seen coming from the
open test port / end of hose.
6 Connect IUT to open test port / end of hose.
Operate Priming Pump until pressure is indicated (max 10
bar)
Operate Volume Adjuster & Refill Valve to achieve required
pressure. (See Section 2.3.5)
Note: If priming a long hose keep the open end of hose vertical to
reduce trapping air.