User Manual Druck DPI612
Page 53 of 105 [EN] English – 109M4017 Revision -
Procedure
1 2 3 4
Step
Procedure
1
If applicable, release the pressure (Section 2.3.6) and remove
the device (Section 2.1.1).
2 Remove the hydraulic pressure release valve.
3 Wind the volume adjuster wheel fully clockwise; this moves
the fluid out of the pressure mechanism
4
Put a container below the instrument then tilt the instrument
up until all the fluid has come out. To discard the hydraulic
fluid, obey all the local health and safety procedures.
Note: Fluid comes out of the test port and the connection for the
pressure release valve.
5
To flush out fluids that contain unwanted material, refill the
system and repeat steps 3 and 4.
Note: To prevent contamination use only one type of hydraulic fluid
in the instrument.