Page 88 of 105 [EN] English – 109M4017 Revision -
Before starting the calibration, check the time and date on the
instrument are correct.
To do a calibration on a measure or source function, use the
advanced menu option.
Select:
Enter the calibration PIN: 4321
Select the button.
Select:
Then select a function and start the calibration:
1. Select channel
2. Select function
3. Select range (if applicable)
4. Follow on-screen instructions
When the calibration is complete, set the next calibration date.
PROCEDURES: CURRENT (measure) 5.5
Do the procedure as follows:
1. Connect the applicable calibration equipment (See Table 5-1)
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3. Use the calibration menu (See Section 5.4) to do a three-point
calibration (-FS, Zero and +FS) for each range: 20 mA and 55 mA.
4. Check the calibration is correct.
Select the applicable Current (measure) function.
DASHBOARD >> SETTINGS
>> ADVANCED
PERFORM CALIBRATION