IRECTION 5813707-100, REVISION 2 VENUE GO™ SERVICE MANUAL
3-50 Chapter 3 - System Setup
PRELIMINARY
3-8-6 Query/Retrieve (Q/R) Setup
3-8-6-1 Overview
The Query/Retrieve function makes it possible to search for and retrieve DICOM data from a DICOM
server for further analysis on the Venue Go™ .
NOTE: You may have to set up Venue Go™ as a destination on the server.
3-8-6-2 Query/Retrieve Setup on the Venue Go™
1) Log on as adm.
2) Select Connectivity > Dataflow
3) In the Dataflow menu, set the Dicom QueryRetrieve slider to On.
4.) From the Dataflow menu, tap to select Dicom QueryRetrieve and to display its properties.
Figure 3-31 Select Dicom QueryRetrieve