D
IRECTION 5813707-100, REVISION 2 VENUE GO™ SERVICE MANUAL
3-52 Chapter 3 - System Setup
PRELIMINARY
2) Select the correct tag from the Select Tag pull-down menu.
3) If needed, type in the value.
4) Select Update List.
5) Select OK to close the Search Criterias window.
3-8-6-3 Query/Retrieve Verification
Figure 3-34 DICOM Query/Retrieve Properties
Follow the steps below to do a first check (TCP-IP Ping) of the connection:
1) If not already selected, select DICOM QR to display the Dicom QR properties.
2) Select the Check button to ping the server.
NOTE: The Check button checks if the remote server is accessible (ping). This isn’t a DICOM Echo
(DICOM ping), so it doesn’t check AE title or port number.
3) If the network connection to the server is OK, it will be illustrated by a green symbol .
A red symbol indicates that the network connection is failing.
Typical causes:
Network cable not connected.
Configuration error(s).
4) When ready, select OK to close the DICOM Query/Retrieve properties and save changes.
Follow the steps below to check the DICOM Query/Retrieve dataflow: