B30 Patient Monitor
11-18
Other adjustable features
NOTE: This section describes the rest of the adjustable features related to ECG
measurement. To adjust the following features, you need a password. If you wish to
adjust the settings of the features, we recommend that you contact the person
responsible for the entire configuration.
ECG printout type
See “Changing printer settings” on page 5-4.
ECG waveform sweep speed
To change the waveform sweep speed, Monitor Setup - Sweep Speeds. Select
Hemodynamics and adjust the value.
Checklist
Check that:
Electrode gel is moist.
Electrodes have good skin contact.
Electrodes are positioned correctly.
Correct leadwire set is selected.
The trunk cable is connected properly.
Leadwire set is properly connected to the trunk cable.
Correct leadwire type is selected in the ECG Setup menu.
Pacemaker selection in the ECG Setup menu is Show when a pacemaker is used.
ECG is selected for screen through Monitor Setup - Screen Setup.