GE HEALTHCARE OPTIMA XR200AMX/XR220AMX
D
IRECTION 5336113-1EN, REVISION 10 INSTALLATION MANUAL
Chapter 1 Safety Page 25
8. Disconnect the two green battery power connectors J3 and J4 and the two sense connectors
J13 and J14 from the Cricket board. Apply a locking device, then apply your personal red lock
and tag. See Figure 1-5.
Figure 1-5 Battery connectors J3/J4/J13/J14 and locking device
Item Description
1 Battery power connectors (J3, J4)
2 Battery sense connectors (J13, J14)
1
2