GE HEALTHCARE OPTIMA XR200AMX/XR220AMX
D
IRECTION 5336113-1EN, REVISION 10 INSTALLATION MANUAL
Chapter 3 Installation Page 53
Figure 3-22 Detector LEDs
3.6.4 Connecting the backup tether
1. Unpack and inspect the tether. Verify that there is no damage.
2. Remove the battery, if present.
3. Connect the backup tether to the detector. Figure 3-23.
Figure 3-23 Connecting the tether to the detector
Battery LED status Meaning
Green, green Battery more than 75% remaining
Off, green Battery remaining between 25% and 75%
Off, yellow Battery remaining between 10% and 25%
Off, red Battery less than 10% remaining
Off, off Detector in sleep mode