Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5 pim 2337741–100
36
Table 18 – Electromagnetic Immunity IEC 60601–1–2
The Innova 2000 system is suitable for use in the specified electromagnetic environment. The
purchaser or user of the Innova 2000 system should assure that it is used in an electromagnetic
environment as described below:
Immunity Test IEC 60601–1–2
Test Level
Compliance
Level
Electromagnetic Environment
Conducted RF
IEC 61000–4–6
Radiated RF
IEC 61000–4–3
3 V
150 kHz to
80 MHz
3 V/m
80 kHz to
800 MHz
3 V/m
800 MHz to
2,5 GHz
[V
1
=
] 3 V
[V
2
=
] 0.3 V
[E
1=
] 3 V/m
[E
2=
] 0.3 V/m
[E
3=
] 3 V/m
[E
4=
] 0.3 V/m
Portable and mobile RF communications equipment are
used no closer to any part of the Innova 2000 system,
including cables, than the recommended separation
distance calculated from the equation appropriate for
the frequency of the transmitter.
At that disturbance amplitude the monitor image might
present some slight flicker. For image quality
improvement, the monitor might be removed from the
electric field source vicinity.
At that disturbance amplitude the system is fully
operational.
At that disturbance amplitude the monitor image might
present some slight flicker. For image quality improve-
ment, the monitor might be removed from the electric
field source vicinity.
At that disturbance amplitude the system is fully
operational.
At that disturbance amplitude the monitor image might
present some slight flicker. For image quality improve-
ment, the monitor might be removed from the electric
field source vicinity.
At that disturbance amplitude level the system is fully
operational.
Note: P is the power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey,* are less than
the compliance level in each frequency range.**
Interference may occur in the vicinity of equipment
marked with the following symbol:
*Field strengths from fixed transmitters, such as base stations for cellular telephones and land mobile radios, amateur
radio, AM and FM radiobroadcast, and TV broadcast cannot be estimated accurately. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be performed. If the measured field
strength exceeds the RF compliance level above, observe the
Innova 2000 system to verify normal operation in each
use location. If abnormal performance is observed, additional measures may be necessary, such as re–orienting or relo-
cating the Innova 2000 system.
**Over the frequency range 150 kHz to 80 MHz, field strengths are less than [V
2
] V/m.
The Recommended Separation Distances are listed in the next table.
Note: These are guidelines. Actual conditions may vary.