Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5 pim 2337741–100
39
Static magnetic field limits:
In order to avoid interference on the Innova 2000 system, static field limits from
the surrounding environment are specified.
Static field is specified less than <1 Gauss in Examination room, and in the
Control Area.
Static field is specified less than <3 Gauss in the Technical Room.
Electrostatic discharges environment & recommendations:
In order to reduce electrostatic discharge interference, install a charge dis-
sipative floor material to avoid electrostatic charge buildup.
The relative humidity shall be at least 30 percent.
The dissipative material shall be connected to the system ground reference,
if applicable.