Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5 pim 2337741–100
47
Illustration 15 – WATER CONDUIT LOCATION
Note: Pipe, junction box and duct or conduit are to be supplied and installed by Customer or
customer’s Contractor.
12” x 12” x 6” BOX
[305mm x 305mm x 152mm]
or similar
24” x 24” x 12” BOX
[610mm x 610mm x 305mm]
or similar
WATER CABLE TROUGH
OR CONDUIT
1” [25mm] I.D. CONDUITS
FOR WATER LINES (FOUR REQ’D)
6” [152mm] I.D. PIPE or CONDUIT
(FLOOR)
LCA BASEPLATE
ISOCENTER
4.5”
[114mm]
0.9”
[23mm]
ELECTRICAL CABLE
TROUGH OR CONDUIT
5.0”
[127mm]
3.6”
[91mm]
WATER LINES
THRU–FLOOR FITTING
(FLOOR)
TABLE SIDE
4.0”
[102mm]
9” or 225 mm (1)
OPENING THRU FLOOR.
ABOVE GRADE
I.D. = Internal Diameter.
Suitable External diameter.
(1) = 9” in US Standard
225 mm in metric Standard.
FLOOR
Note: Flush floor interface fitting is part of GEMS installation kit 2286398 and is installed by
Customer or customer’s Contractor.
ON GRADE
LC+
Isocenter
LC+
Baseplate
Cable conduit
(flush floor
interface fitting)
4 water pipes:
2 between MAXIRAY 150
2 between Revolution Digital Detector
and chillers
about 3” (76 mm)
internal diameter for
water sheath
Water pipe
conduit
separated from
electrical conduit
3” DIA.
(76 mm)
Flush floor
duct
15” (38 cm)
bend radius
minimum