NOTE: (3): In use temperature limits specify the range where the system shall work. Operating
outside these limits could occur severe performance and reliability issues.
Recommended temperature limits specify the range where it is recommended to adjust
air conditioning control in order to warranty current operations inside the in use range.
Relative Humidity and Temperature: Refer to Table 3-1. To obtain relative humidity and
temperature requirements for components not specified in Table 3-1, refer to the appropriate
component Pre–Installation Manual listed in Chapter 2, Section 3, Basic Innova System
Compatibility.
Altitude and Atmospheric Pressure: Refer to Table 3-2. To obtain altitude and atmospheric
pressure requirements for components not specified in Table 3-2, refer to the appropriate
component Pre–Installation Manual listed in Chapter 2, Section 3, Basic Innova System
Compatibility.
Table 3-2:
INSTALLATION ROOM
OF PRODUCT OR COM‐
PONENT
ALTITUDE (meters)
ATMOSPHERIC PRESSURE (kPa)
IN-USE
STORAGE / TRANS‐
PORT
IN-USE
STORAGE / TRANSPORT
MIN
MAX
MIN
MAX
MIN
MAX
MIN
MAX
Examination room
0
3000
-30
3000
70
106
70
106
Technical room (See No‐
tice below
0
3000
-30
3000
70
106
50
106
Control room
0
3000
0
3000
N/A
N/A
N/A
N/A
The chiller is able to dissipate maximum continuously power at 3000 meters in an
ambient temperature up to 20°C, for this altitude the technical room temperature
shall not exceed 20°C.
In some cases condensation occurs and water drops from outlets and pipes of the
air conditioner in the technical room.
Therefore, it is critical to install the cabinets where there is no risk of flood from the
air conditioner.
Ensure the air outlet positions in the exam room is not in area closed to monitor
suspension carriage/rail.
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
Chapter 3 Physical Requirements of Room 55