2.2 Room Layout Drawings
2.2.1 Patient Room Layout
2.2.1.1 Room Layout for Innova System (examples of room layouts)
WARNING
LOCATION IN TECHNICAL ROOM FOR ELECTRICAL CABINETS IS
MANDATORY.
THE ELECTRONIC CABINETS (C1, C2, OPTIONAL LD CABINET, COOLIX
4100 CHILLER, DETECTOR CHILLER AND FLUORO UPS WHEN
INSTALLED ) INCLUDE FANS THAT ARE CREATING AIR–CIRCULATION OF
PULSED–AIR. WHEN THIS PULSED AIR IS IN AN ENVIRONMENT THAT MAY
CONTAIN AIRBORNE PATHOGENS LIKE AN EXAM ROOM/CONTROL
ROOM, THERE IS A RISK OF TRANSMISSION OF THESE AIRBORNE
PATHOGENS FROM PATIENTS TO OTHER PATIENTS OR CLINICAL
PERSONNEL (NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE
INSTALLED IN A ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM,
I.E., TECHNICAL ROOM.
Legend of Illustration 2-53, Illustration 2-54, Illustration 2-55 and Illustration 2-56.
1.
Gantry
2. Patient table
3. XT monitor suspension
4. Monitor frame
5. Injector on Pedestal
NOTE: The phone outlet must be located less than 1 meter (3 feet) from the C1 Cabinet (3).
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2