3 Room Structural Requirements
3.1 General Policy
GE Healthcare’s Customer is responsible for the structural analysis and mounting of the base
plates. If GE Healthcare is forced to mount the base plate, the LCT must hire a structural
engineer to design and approve the mounting method and provide GE Healthcare with an
engineering report.
The floor level cannot exceed a general levelness of 5 mm (0.2 in) for any 2 meters (79 in).
NOTICE
The floor slabs on which the equipment is to be installed must have a levelness
of 1 mm (0.04 in) per meter (40 in). Position of baseplates and table basement
depends on the type of installation. The two types of installation are given below.
Illustration 2-65: Types of baseplate installation
NOTICE
The Table baseplate or the System baseplate are mandatory to install the Tilting
table.
The Tilting table must never be installed on grade.
NOTICE
Augmented Calibration requires that the patient table and gantry be installed
following Pre-installation instructions. Any deviation may result in a non-
augmented calibration.
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
Chapter 2 Equipment Requirements 129