Illustration 2-69: Table floor mounting layout
NOTICE
Prepare the floor such that the table base plate will be flush with the floor finish
surface, taking into account the thickness of the floor finish material.
For alternative table bolts or seismic area, refer to template drawing shown in Illustration
Innova
LC Positioner And Table Floor Mounting Template
contained in Mounting Requirements.
3.2.3 Hole dimension and preferred location in concrete floor
In the examination room, the Innova LC Positioner is not placed on a computer floor but directly
put on concrete floor, the location of the cable access needs to be carefully planned.
Otherwise, if the cable run is located under the concrete floor, the cables will have to come
through the floor and in this case you will need two holes, one for the LC Positioner and the
other for the patient table.
The diameter of both holes must be the same 225 mm or 9 in.
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
132 3 Room Structural Requirements