Chapter 4 Environmental Requirements
1 Relative Humidity and Temperature
NOTICE
Avoid extremes in temperatures
Innova system room climate requirements – relative humidity and temperature (individual
products or components are classed by their installation area for In use see -Table 4-1)
Table 4-1:
INSTALLATION
ROOM OF
PRODUCT OR
COMPONENT
RELATIVE HUMIDITY (NON- CON‐
DENSING)
TEMPERATURE
IN-USE IN-USE (See Note (2a)) RECOMMENDED (See Note (2b))
MIN MAX MIN MAX MIN MAX
Examination
room
30% 70% +15°C +59°F +32°C +90°F
Design for Patient/ Operator Com‐
fort
Technical room
(See Note (1))
30% 75% +10°C +50°F +30°C +86°F +13°C +55°F +25°C +77°F
Technical room
with Fluoro UPS
optional
30% 75% +20°C +68 °F +25°C +77°F +20°C +68°F +25°C +77°F
Control room 30% 75% +15°C +59°F +35°C +95°F +20°C +68°F +25°C +77°F
NOTE: (1): The target temperature (best recommended) is 18°C (64°F).
NOTE: (2a): In use temperature limits specify the range where the system shall work.
Operating outside these limits could occur severe performance and reliability issues.
(2b): Recommended temperature limits specify the range where it is recommended to
adjust air conditioning control in order to warranty current operations inside the in use
range.
Relative Humidity and Temperature: Refer to Table 4-1. To obtain relative humidity and
temperature requirements for components not specified in the table, refer to the appropriate
component Pre–Installation Manual listed in Chapter 2, Basic Innova System Compatibility.
NOTICE
In some cases condensation occurs and water drops from outlets and pipes of
the air conditioner in the technical room.
Therefore, it is critical to install the cabinets where there is no risk of flood from
the air conditioner.
NOTICE
Ensure the air outlet positions in the exam room is not in area closed to monitor
suspension carriage/rail.
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
Chapter 4 Environmental Requirements 187