NOTE: Refer to the UPS vendor Service Operating Manual for more details:
•
UL Model: UPS Option - Operating Manual SG-UL Series - 20 kVA
•
CE Model: UPS Option - Operating Manual SitePro 20 kVA (series 8)
3.3 IVUS Option
Device Location Heat output
PC Control room 335W – 1206 BTU/h
Monitor Control room 125W – 427 BTU/h
Printer Control room 67W – 229 BTU/h (idle)
NOTE: For IVUS Rev 3, refer to
Volcano s5i Imaging System Integration in IGS Systems -
Service Manual.
WARNING
TAKE ANY STEPS NECESSARY TO PROVIDE CORRECT TEMPERATURE
AND DUST FREE ENVIRONMENT FOR CORRECT OPERATION OF PC IN
THE CONTROL ROOM.
3.4 Large Display Subsystem Option
NOTICE
The Large Display Cabinet requires a cooling system to keep ambient
temperature below 35 ºC (95 ºF).
The heat produced by the Large Display Cabinet is transferred to the environment by its
ventilation. Cooling air enters the Large Display cabinet through the air inlet (grids) located at
the front and exhausted through the outlet on the rear & from underneath the Large Display
Cabinet. A suitable ventilation or cooling system must be installed to extract the heat from the
Large Display Cabinet room
Ventilation holes must not be covered or closed; otherwise the heat generated in the device
cannot be dissipated sufficiently.
Avoid dusty environments:
The Large Display Monitor has been designed for use in the clean environment of medical
diagnostics. The Large Display Monitor dissipates heat through the openings at the rear. Dust
from dirty environments can penetrate into the display through these openings. In the extreme
case, deposits are possible which become evident as dark spots in a white picture and which
can result in deterioration of the luminance. Protect the display from dust, e.g. during building
measures at the installation location, and use the original packaging for transport.
NOTICE
Do not put anything on the top of the Large Display cabinet.
If the Large Display Cabinet runs in a dusty environment, we recommend strongly installing
filters on the air inlet of the Large Display Cabinet room. In this case it should be considered that
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
Chapter 4 Environmental Requirements 193