2.4.4.2 Run Interconnection cables through the Conduit
2.4.4.2.1 For IVUS Rev 2 option
Use the IVUS Rev 2 PIM (#806365002) as reference to the procedures below.
Unpack , and check the completeness of the Preinstallation kit :
1 pcs - Custom PIM cable 30 m (98 ft, 5 in) - Volcano ID :806452001
1 pcs - CAT5 cable 30 m (98 ft, 5 in) - Volcano ID :807055001
1 pcs - Custom Twin axial cable - Volcano ID :806380001
1 pcs - Ground Cable, 30 m (98 ft, 5 in) s5i – Volcano ID :806889001
1 pcs - IVUS Rev 2 Preinstallation Manual - Volcano ID :806365002
Install 806889001 Ground Cable through the Control Booth conduit (see note)
Install the cables through the IVUS conduit from the Control Room end :
806452001 PIM cable, round shaped connector end to the Procedures Room
807055001 - Ethernet cable (1 pcs)
Install the optional cables through the IVUS conduit from the Control Room end :
806380001 - ECG cable in case, if NON Mac-Lab ECG will be installed
808631002 – FFR cable option - 15m
808632002 – FFR cable option - 27m
NOTE:
For the ground cable, see the IVUS Rev 2 service manual for special
instructions concerning cable running and cutting .
NOTE:
If the installation has the MaC-Lab ECG in the configuration, the ECG cable
shall not be run through the conduit, it will be used in the Control Room to
connect Mac-Lab and IVUS PCs.
NOTE:
You should consider running the ECG cable through the conduit in case, the
customer requires interconnecting to a NON Mac-Lab ECG at the table base.
In this case you shall hide the unconnected cable in the conduit for future
expansion options.
NOTE: Same applies to the 2nd Ethernet cable, if a later extension of the IVUS
remote controls expected.
Ensure sufficient cable length on the Procedures Room end of the cables according to the
table below
Cable Free length from the conduit exit
IVUS PIM cable * 2.4 m (8 ’)
FFR cables for Pimette ** 0m (0’)
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
228 2 System Cable Information