Illustration 2-64:
CAUTION
The service area dimensions shown above are minimum requirements. Service
areas must comply with local regulations if more stringent.
2.3.8 Layout Constraints
NOTICE
The X-Ray tube Chiller elevation cannot exceed 3 meters (9.84 feet) in height.
NOTICE
Minimum distance between Digital Detector and UPS is 3 m (118 inches) (Fluoro
UPS option).
NOTICE
The distance of the Detector chiller cannot exceed 3 m (1 floor) in height below
the detector.
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
128 2 Room Layouts