GE Innova IGS 540 - 3.2.4 Water Pipe Requirements

GE Innova IGS 540
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 2-7: D distance
ANGIO CARDIO NEURO
Omega IV Compact NA 1395 mm (54.9 in) NA
Omega V Long 1278 mm (50.3 in) 1395 mm (54.9 in) 1395 mm (54,9 in)
Omega V non motorized Long 1278 mm (50.3 in) 1395 mm (54.9 in) 1395 mm (54,9 in)
Tilting table 1278 mm (50.3 in) 1395 mm (54.9 in) 1595 mm (62.8 in)
NOTICE
Due to the plastic bushing used in the USA to protect cables from the sharp
edges of conduits it is necessary to place the cable conduit inside the table cable
access opening but the height of the outcoming conduit plus bushing is limited to
1/2 in (12.7 mm).
NOTE: Refer to table
Chemical anchors Pull out efforts and recommendations
in Mounting
Requirements for pull out effort on each fixation bolts.
3.2.4 Water Pipe Requirements
An Innova LC System uses a Performix 160 A X–Ray tube with a recirculating chiller.
Two water hoses are supplied to allow water circulation between Innova LC Positioner and
chiller.
In some countries, it is forbidden to run electrical cables and water pipes in the same
conduit. In this case, two separate conduits are required. But then a problem arises at the
level of the Innova LC Positioner entrance.
Depending on the method used, one of the two options shown hereafter (Illustrations
Illustration 2-71 and Illustration 2-72) must be used.
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
134 3 Room Structural Requirements