4 Mounting Data, Including Seismic
4.1 Seismic areas
4.1.1 Floors
Chiller: The seismic kit for the Chiller 4100 (SMC) is included with the chiller when shipped.
LD cabinet: the seismic kit to fasten the LD cabinet to the wall is included with the cabinet when
shipped.
In Seismic areas all cabinets must be anchored to the floor. See Basic Innova System
Compatibility for referential documents. C1 & C2 Cabinets: Wall support 2284542-2.
Every sub–system is delivered on site with its proper seismic kit.
Monitor Flat Panel Seismic Kit: 2353317
VCIM seismic kit: 2365510.
Anti–seismic means be installed before opening the system for normal use.
4.1.2 Walls
NOTE: For Fluoro UPS option (US version only), the Fluoro UPS supplier will deliver
anchoring brackets (see Illustration 2-86). The bolts will be provided by the customer.
Consider local seismic codes when planning cabinet mounting. Consult seismic expert to
determine which mounting method is appropriate for the seismic region. Certain seismic regions
require additional reinforcement in walls. See Basic Innova System Compatability for referential
documents.
Illustration 2-86:
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
164 4 Mounting Data, Including Seismic