However, when installed and operated as specified herein, the system will maintain its essential
performance by continuing to acquire, display, and store diagnostic quality images safely.
For example, a 1W mobile phone (800 MHz to 2.5 GHz carrier frequency) shall be put 2.3
meters (see (*) Table 3-4) apart from the Innova system (in order to avoid images interferences
risks).
2.5 Use Limitation
WARNING
THE USE OF ACCESSORIES, TRANSDUCERS, AND CABLES OTHER THAN
THOSE SPECIFIED MAY RESULT IN DEGRADED ELECTROMAGNETIC
COMPATIBILITY OF THE INNOVA SYSTEM.
In case of an Innova system, the Gantry (digital detector) shall be apart 1 meter from the
generator cabinet, and 1 meter apart from the analog (CRT) monitors. These distances
specifications shall minimize the low frequency magnetic field interference risk.
Other electrical equipment may disturb and interfere with these Innova components. The control
of the clearing distances from the noise sources is recommended from the HF electrosurgery
generator, power supplies converters from nearby monitors or from other close electrical
equipment). Refer to respective device manufacturers instructions & recommendations in such
cases.
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
184 2 EMI Consideration