Cable # From To Type Characteristics Comments
12 Mains LDM UPS
Phase, neutral and
ground
UL: AWG10 (5.26 mm²)
CE: AWG12 (2.5 mm²)
1.2.3 LOTO devices provided by the installer
1.2.3.1 Core system
A wall circuit breaker or equivalent device with LOTO capability must be installed on the mains
line to the PBD. This device must be compatible with the power input specifications of the
system. The customer is responsible for the procurement, delivery and installation of this
breaker.
1.2.3.2 LDM option
A wall circuit breaker or equivalent device with LOTO capability must be installed on the mains
line to the LDM UPS. The rating of this device shall be 30A for UL and 16A for CE
configurations. The customer is responsible for the procurement, delivery and installation of this
breaker.
1.2.4 System interconnections
1.2.4.1 Emergency power off
The PDB (in technical room) is provided with an EPO button on its front panel. The customer is
given the possibility to install additional EPO buttons (for instance in the exam room and in the
technical room). The PDB is designed for 2 additional EPO buttons; these buttons shall be
“Push to activate - Push to release” type, 2 contacts Normally Closed, compatible with 24 VDC.
The UL PDB is provided with 2 additional EPO buttons, the CE PDB is provided without
additional EPO buttons.
The customer is responsible for the procurement, delivery and installation of the cables for
these additional EPO buttons. The max length shall be 24 m, the recommended diameter is
AWG14/2 mm².
Protect the Emergency Stop from accidental actuation.
1.2.4.2 X-Ray ON lights
NOTICE
THE X-RAY ON LAMP MUST BE INSTALLED IN THE EXAM ROOM IN
CONFORMITY TO THE STANDARD IEC/EN 60601-2-43. THE X-RAY ON
LAMP SHALL BE VISIBLE BY THE OPERATOR IN ALL THE LOCATIONS
DEFINED FOR THE PERSONNEL WHO MAY RECEIVE SCATTERED
RADIATION.
The system is designed to provide power for X-ray ON lights. The customer is responsible for
the procurement, delivery and installation of the cables and the X-ray ON lights.
UL system:
The terminals 2 & 3 from the Positioner Bulkhead of the C2 cabinet shall be connected to PDB
TB1 terminals 24 & 25 (cable provided by the installer, maximum length 24 m, minimum
diameter AWG16). The lamp shall be 120V ~ 1A max, it shall be connected to PDB TB1
terminals 25 & 26, with cable AWG16 minimum
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
202 1 Power Requirements