NOTE: Never connect the Physio module to the ground.
The Physio cable can run in the same conduit as the Bolus cable. In this case, it is
required to have a conduit between the table and the physio gases box.
If no conduit available between rear of table and Control Room (no Remote TSUI, no
IVUS, no MacLab…), need to define proper cable routing or create new conduit as
per PIM requirements.
If there is no physio gases box behind the table in the lay out, find a local solution to
hide the hole in the floor and the cable exit.
On some sites there is only one conduit between Exam room and Control room for the
IVUS & MacLab.
2.4.4 IVUS S5I Option
2.4.4.1 Floor Conduits
NOTE: (For IVUS Rev 3 option) refer to Volcano s5i Imaging System Integration in IGS
Systems - Service Manual.
A 3 inches and max. 25 m long conduit between Control and Patient Rooms shall be prepared
below the floor for the S5I GE Innova interconnection cables.
•
Locate the position of the S5I GE components in the Control Room, according to the
workflow requirement of the operators.
•
Establish a 76 mm (3”) conduit below floor from the S5I GE CPU location in the Control
Room, to the foot end area of the Patient Table in the Procedures Room:
○
Locate the Procedures Room exit of the conduit close to the centerline of the Patient
Table
○
Locate it as close to the Patient Table base as possible, considering :
■
Room for access to the table for service purposes
■
Limitation from other cables connected to the table
■
Location of the table plate
■
Location of Mac-Lab conduit
■
Location of Physio or Med Gases connections
■
The S5I GE cables are moving cables, they have to allow the Patient Table motions
(Panning, Lifting, Tilting and Rotating)
•
Use similar solutions (i.e. protection against cleaning fluids, etc. ) as it is used for the Mac-
Lab conduit
•
See also guidance on Typical Layout drawing 4-58f sheets E1 and S2, showing Mac-Lab
and P/M Gas location recommendations.
Estimated labour time: 1 hour
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
Chapter 5 Electrical Requirements 227