Cable Free length from the conduit exit
Control Station cable 2.1 m (7 ’)
Joystick cable 2.7 m (9 ’)
ECG cable Depends on the ECG location
* Consider the 12’ length of the PIMr cable, if PIMr exists.
** Pimette has 3 m cable.
NOTE:
The incorrect cable length might cause reliability issues during the operation !
Make sure, that the out of conduit parts of the cables with the connectors are properly protected
during the rest of the installation
2.4.4.2.2 For S5I GE Rev 2+ option
Use the S5I GE Rev 2+ PIM (#806365026) as reference to the procedures below.
•
Unpack , and check the completeness of the Preinstallation kit:
○
1 pcs - Custom PIM cable 27 m (88.2 ft) - Volcano ID: 806452008
○
1 pcs - CAT5 cable 30 m (98 ft, 5 in) - Volcano ID: 807055001
○
1 pcs - Custom Twin axial cable - Volcano ID: 806380001
○
1 pcs - Ground Cable, 30 m (98 ft, 5 in) s5i – Volcano ID: 806889001
○
1 pcs - S5I GE Rev 2+ Preinstallation Manual - Volcano ID: 806365007
○
1pcs – FFR cable 27 m - Volcano ID: 808632001
•
Install 806889001 Ground Cable through the Control Booth conduit (see note)
•
Install the cables through the S5I GE conduit from the Control Room end :
○
806452008 PIM cable, round shaped connector end to the Procedures Room
○
807055001 - Ethernet cable (1 pcs)
○
808632001 – FFR cable - 27 m
•
Install the optional cables through the S5I GE conduit from the Control Room end:
806380001 - ECG cable in case, if NON Mac-Lab ECG will be installed
NOTE:
For the ground cable, see the S5I GE Rev 2+ service manual for special
instructions concerning cable running and cutting .
NOTE: If the installation has the Mac-Lab ECG in the configuration, the ECG cable shall
not be run through the conduit, it will be used in the Control Room to connect Mac-
Lab and S5I GE PCs.
NOTE: You should consider running the ECG cable through the conduit in case, the
customer requires interconnecting to a NON Mac-Lab ECG at the table base.
In this case you shall hide the unconnected cable in the conduit for future
expansion options.
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
Chapter 5 Electrical Requirements 229