Chapter 2 Equipment Requirements
1 System Components
1.1 Presentation of the 3 Rooms
1.1.1 Examination room
Gantry
Omega / Tilting table
Performix X–Ray Tube Assembly
Collimator
Innova Digital Detector
WARNING
THE ELECTRONIC CABINETS (C1, C2, OPTIONAL LD CABINET, COOLIX
4100 CHILLER, DETECTOR CONDITIONER AND FLUORO UPS WHEN
INSTALLED) INCLUDE FANS THAT ARE CREATING AIR–CIRCULATION OF
PULSED–AIR. WHEN THIS PULSED AIR IS IN AN ENVIRONMENT THAT MAY
CONTAIN AIRBORNE PATHOGENS LIKE AN EXAM ROOM/CONTROL
ROOM, THERE IS A RISK OF TRANSMISSION OF THESE AIRBORNE
PATHOGENS FROM PATIENTS TO OTHER PATIENTS OR CLINICAL
PERSONNEL (NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE
INSTALLED IN A ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM,
I.E., TECHNICAL ROOM.
1.1.2 Technical room
C2 Cabinet
C1 Cabinet
Coolix 4100 chiller
Detector Conditioner
Fluoro UPS (option) (1 Cabinet)
Fluoro UPS IF box (option)
Large Display Cabinet and UPS (for Large Display Option)
1.1.3 Control room
VCIM
DL Flat Panel
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
Chapter 2 Equipment Requirements 47