4.3 Heat Output
Equipment Heat Output Tables
Refer to Table 4-1 on page 131. To obtain heat output information for components not specified in
Table 4-1 on page 131, refer to the appropriate component Pre-Installation Manual listed in
2.1.5 System Compatibility on page 71.
Table 4-1
HEAT OUTPUT
Stand by
Moderate Use (8
cases / a 10 hour
day)
Typical Use (13
cases / a 10 hour
day)
Maximum Use
(maximum peak
power during ex-
am)
Room Core System kW BTU/hr kW BTU/hr kW BTU/hr kW BTU/hr
Exam
Room
Innova Frontal/Later-
al positioner and ta-
ble
0.61 2076 0.75 2540 1.21 4128 1.62 5517
Ctrl
Room
DL user area with 1
TFT monitor
0.16 546 0.16 546 0.16 546 0.16 546
3 B&W flat monitors 0.25 859 0.25 859 0.25 859 0.25 859
Ctrl Room Total 0.41 1405 0.41 1405 0.41 1405 0.41 1405
Tech
Room
C1 Frontal Cabinet 0.71 2421 0.99 3389 1.29 4412 1.59 5435
C1 Lateral Cabinet 0.31 1057 0.56 1923 0.86 2946 1.16 3969
C2 Cabinet 0.29 989 0.83 2813 1.34 4571 1.81 6171
Coolix 4100 chiller
Frontal & Lateral @
50Hz (1) (2)
1.55 5288.6 2.1 7165 2.72 9280.64 5.5 18766
Coolix 4100 chiller
Frontal & Lateral @
60Hz (1) (2)
2.32 7915.84 2.8 9553 3.44 11737.2
8
6.3 21495.6
Chiller autotrans-
former @ 50Hz
0.04 136.48 0.048 164 0.045 153.54 0.065 221.78
Chiller autotrans-
former @ 60Hz
0.06 204.72 0.064 218 0.07 238.84 0.09 307.08
Detector conditioner
Frontal
0.21 709 0.21 709 0.21 709 0.21 709
Detector conditioner
Lateral
0.21 709 0.21 709 0.21 709 0.21 709
Environmental Requirements
5750182-1EN 3
Innova
TM
IGS 6
131