Classification category Equipment classification
Degree of protection against
harmful ingress of water
Ordinary equipment (enclosed equipment without protection against ingress
of water), except footswitch which is a watertight device (protected against
the effects of submersion, IPX8).
Method(s) of sterilization or disin-
fection recommended by the
manufacturer
• Sterilization: not applicable
• Disinfection: refer to Operator Manual (Chapter Safety and Regulatory,
section Disinfection), Recommended disinfecting agents.
Degree of safety of application in
the presence of a flammable an-
esthetic mixture with air or with
oxygen or nitrous oxide
Equipment not suitable for use in the presence of a flammable anesthetic
mixture with air or with oxygen or nitrous oxide
Mode of operation Continuous operation with intermittent loading
NOTICE
The system can only be installed in an anesthetizing location if that location is
classified as Other Than Hazardous as per NFPA 70 clause 517.60
NOTICE
The product is not classified as AP, APG (Equipment not suitable for use in the
presence of a flammable anaesthetic mixture with air or with oxygen or nitrous
oxide).
1.2.3 Pre-Installation Checklist
Refer to the document Global Site Readiness Checklist DI - DOC1809666 for standard HPM requirements
on Room preparation for Vascular Systems installation.
See also the specific preparation requirements for IGS Systems installation given in sections 3, 4 and 5 of
the Tab "Installation Prerequisites" in document IGS System Installation Prerequisites - DOC2024755.
NOTE
DOC1809666 and DOC2024755 are available from MyWorkshop.