Large Display option requirements:
The Large Display option must be powered through a wall circuit breaker or equivalent device with LOTO
capability. This circuit breaker must have a 30 amps current rating. Procurement, delivery and
installation of this device is customer’s responsibility. The customer shall provide the power supply cables
from this circuit breaker to the LDM UPS, and the ground cable from the main power to the LDM cabinet.
Table 5-6 System FEEDER from hospital
Power supply voltage
480 V
Panel (PDB) to C1 run in
m (ft)
8 (26) 16 (52) 24 (79)
Feeder run
length:
10 m
(30 ft) 2/0 2/0 2/0
WARNING
RECOMMENDED FEEDER CABLE GAUGE :
FOR PHASES, NEUTRAL & GROUND PDB
INPUT TERMINALS ACCEPT 2/0 AND 3/0
AWG. BIGGER CABLES NEEDS A SIZE RE-
DUCTION BEFORE PDB ENTRY.
20 m (70 ft) 2/0 2/0 2/0
30 m (100 ft) 2/0 2/0 3/0
40 m (130 ft) 2/0 2/0 250
50 m (160 ft) 2/0 3/0 300
60 m (200 ft) 3/0 4/0 400
80 m (260 ft) 4/0 350 600
100 m (330 ft) 300 500
see
Note
120 m (390 ft) 400 600
see
Note
Power supply voltage
380 V 400 V 415 V
Panel (PDB) to C1 cabinet run
8 m 16 m 24 m 8 m 16 m 24 m 8 m 16 m 24 m
Feeder run
length:
10 m
35 35 35 35 35 35 35 35 35
20 m 35 35 70 35 35 50 35 35 50
30 m 50 70 95 35 50 70 35 50 70
40 m 50 70 120 50 70 95 50 70 95
50 m 70 95 185 70 95 120 70 95 120
60 m 95 120 240 70 95 185 70 95 150
Electrical Requirements
5750182-1EN 3