Chapter 6. Communication Requirements
6.1 Network Requirements
General Information
The system is provided with a firewall unit, that allows connection to the hospital network for pushing the
DICOM images or for service remote access (InSite). This firewall is compatible with 10/100/1000 (Gigabit
Ethernet) networks.
The C-FRT Cabinet provides an Ethernet RJ45 plug, the hospital is responsible for providing the Ethernet
cable between the system and the hospital network.
For complete descriptions of these connectivity solutions, please refer to the Broadband Solutions
catalogue available through your local GEHC sales and service representative.
Connectivity Process and pre-installations checklists are available in the Broadband Connectivity PIM
available through your local GEHC sales and service representative.
InSite Connection
InSite requires a VPN connection to the system. To create this connection, the system IP and site IT
contact information, should be given to your local GEHC sale and service representative before the
system installation begins. Once submitted, a member of the GEHC Broadband Solutions team will
contact the site IT to set up the VPN connection.
The SupportCentral links where information from the InSite Connectivity Team or Insite Request Form can
be found are:
Americas: http://supportcentral.ge.com/products/sup_products.asp?prod_id=73661
Asia: http://supportcentral.ge.com/products/sup_products.asp?prod_id=19181
Europe: http://supportcentral.ge.com/products/sup_products.asp?prod_id=24026
IP Adresses
IP addresses for DL and AW PCs have to be requested to the Hospital IT at the time of pre-install to not
delay the installation.
New requirements related to the Privacy and Security configurations also apply with the new system
software generation. Refer to 6.2 Privacy and Security Configuration on page 181.
6.2 Privacy and Security Configuration
The new Privacy and Security features available with the System require to be configured according to
the security policy requested by the hospital.
To ensure the installation is successful and is not delayed because of missing information, it is required to
gather all needed information as part of the pre-install process.