2.2 Room Layouts
2.2.1 Room Dimension Requirements
For Room size dimensions, refer to 2.2.2.1 Patient Room Layout on page 72.
For additional details, refer to 2.2.3 Room Layout Considerations on page 80.
2.2.2 Room Layout Drawings
2.2.2.1 Patient Room Layout
WARNING
LOCATION IN TECHNICAL ROOM FOR ELECTRICAL CABINETS IS MANDATORY. THE
ELECTRONIC CABINETS (C1, C2, OPTIONAL LD CABINET, COOLIX 4100 CHILLER,
DETECTOR CONDITIONER AND FLUORO UPS WHEN INSTALLED ) INCLUDE FANS
THAT ARE CREATING AIR-CIRCULATION OF PULSED-AIR. WHEN THIS PULSED AIR
IS IN AN ENVIRONMENT THAT MAY CONTAIN AIRBORNE PATHOGENS LIKE AN
EXAM ROOM/CONTROL ROOM, THERE IS A RISK OF TRANSMISSION OF THESE
AIRBORNE PATHOGENS FROM PATIENTS TO OTHER PATIENTS OR CLINICAL
PERSONNEL (NOSOCOMIAL DISEASES).
TO REDUCE THIS RISK, THE ELECTRONIC CABINETS MUST BE INSTALLED IN A
ROOM SEPARATED FROM EXAM ROOM/CONTROL ROOM, I.E., TECHNICAL ROOM.
Table 2-8 Exam room dimensions
Room Dimensions Length x Width Ceiling Height
Recommended:
11570 mm x 8000 mm (38 ft 0 in x 26 ft 2
in)
2845 mm ± 5 mm (9 ft 4 in ± 0.2 in) is man-
datory ceiling height
Minimum: 6900 mm x 4400 mm (22 ft 8 in x 14 ft 5 in)
Equipment Requirements
72
Innova
TM
IGS 6
5750182-1EN 3