Illustration 1-1:
2. Draw a line on the floor going through the center of the hole and extending in the direction the
table is to be installed. This is the isocenter line. It passes through the gantry isocenter and
the table pivot point.
Single Plane and Biplane Innova Systems Pre–Installation Kit Installation Procedures
GE Healthcare Direction 2290880-12-1EN, Revision 1
18 1 PIST0002 - Prepare Floor - LC Base Plate and Omega/Innova IQ Table