MA-041 DGA 900 Plus - Installation & Commissioning Manual - Rev 1.1 4-Mar-22 Page 40 of 99
7. Use a heat gun to shrink the 3/16 in.
tubing.
8. Then, insert the ¼ in. tubing to
overlap about 15 mm (0.6 in.) over
the metallic braid.
9. Use a heat gun to shrink the ¼ in.
tubing.
10. Strip away about 7 mm (¼ in.) of
insulation from the central conductor.
Twist the strands together.
9.4.1.1.4 Connecting to the Bushing Adaptor
At the Bushing Adaptor end, the coaxial cables must be terminated with two pin
terminals as outlined in Table 9-4.
Table 9-4: Connecting to the Bushing Adaptor
Steps
Visual
1. Select two red pin terminals and the
corresponding hand crimp tool.
2. Insert the central conductor of the
coaxial cable into one pin terminal and
crimp that terminal.
3. Insert the braid shield into the second
pin terminal and crimp that terminal.
4. The terminals must be double crimped
(once on the bare wire and once on
the insulation). Verify visually and by a
pull test on the terminals.