DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
7-30 Section 7-7 - Trouble Shooting Tree
7-7-6 Monitor Troubleshooting
NOTE: Reset the monitor settings to the factory defaults prior to troubleshooting the monitor.
NOTICE
Comments to remember about the 15” monitor.
The monitor is the diagnostic monitor and is connected to AC 110/220 VAC.
The monitor powers down with the console.
Power output from AC Power assy via AC distributor in AC Transbox assy:
When the monitor is not connected to the AC power there is a leakage current which
when measured with a DVM can indicate high voltage.
When a load is applied, the voltage collapses to low voltage if the monitor is not
switched on.