Chapter 8: System Hardware Overview
5495975-1EN Rev.9 8-43
© 2013-2017 General Electric Company. All rights reserved.
Figure 8-29 Front of wallstand receptor with four ion chambers
When the receptor is at or near 0° (vertical), the three main chambers are active and the fourth (lower left)
is disabled (Figure 8-30).
Figure 8-30 Receptor in vertical configuration
When the receptor is at or near 90° (horizontal), the fourth chamber becomes enabled and the top right
chamber becomes disabled (Figure 8-31). On the Image Acquisition screen ion chamber selection, the
“dashed” ion chamber is assigned to position 1.
Fourth ion
chamber