Chapter 10: Image Acquisition
5495975-1EN Rev.9 10-33
© 2013-2017 General Electric Company. All rights reserved.
Figure 10-24 Image Recovery confirmation
During this time, do not remove the active detector from the table or wallstand. If in wireless mode place
the detector within a strong wireless signal area – do not power off. Select the ‘Recover Image’ button to
continue image recovery. Select the ‘Close Exam’ button to stop the recovery process and close the
exam.
CAUTION Images will be lost if the recovery process is stopped.
When the image is recovered, it will appear on the viewer.