Chapter 14: Quality Assurance and Maintenance
5495975-1EN Rev.9 14-6
© 2013-2017 General Electric Company. All rights reserved.
5. From left pane, open Quality Control Tools > Acquisition. (Figure 14-7)
Selecting Acquisition.
Figure 14-7
The Start screen appears (Figure 14-8).
Figure 14-8 Start screen
6. Select the detector receptor to perform QAP.
Click on QAP. The Detector Check will begin.
The Flat-field Tests screen appears (Figure 14-9).