Chapter 14: Quality Assurance and Maintenance
5495975-1EN Rev.9 14-9
© 2013-2017 General Electric Company. All rights reserved.
Figure 14-11 Composite Tests screen
10. Follow the instructions at the bottom of the MTF Tests screen:
a) Remove the flat-field phantom or any other objects from the field of view.
b) Insert the MTF phantom into the grid holder.
c) Prep and Expose the MTF phantom.
The phantom image appears on the monitor (Figure 14-12).
The Image Quality Test Results appear.