Chapter 14: Quality Assurance and Maintenance
5495975-1EN Rev.9 14-10
© 2013-2017 General Electric Company. All rights reserved.
Figure 14-12 MTF phantom image
11. Review the results.
If Pass: QAP is complete. Click [EXIT] to return to Worklist screen.
If Fail: Refer to Chapter14: -
Failed QAP (p.14-11) for more information.
12. Log off and perform a controlled shutdown and start up. Refer to Chapter4: -
System Start Up and
Shutdown (p.4-2) for more information.