Appendix B: Specifications
5495975-1EN Rev.9 B-1
© 2013-2017 General Electric Company. All rights reserved.
Appendix B: Specifications
X-Ray Tube:
Heat Capacity: 350Khu
Focal Spot: 0.6mm/1.25mm (1.3 IEC)
Tube Type: Anode Rotation
Nominal Anode Input Power: 32kW/100kW
Leakage Radiation parameter: 150kV/4mA
Note: For the specific and detailed X-Ray Tube specifications, please refer to the accompanying Docu-
ment of the Tube.
Collimator
Leakage Radiation parameter: 150kV/3mA
Maximum field size [cm] with SID = 100cm 47.8 * 47.8
Storage Conditions:
Environmental Temperature: -20
°C~60°C (-5°C~50°C for Wireless Detector)
Relative Humidity: 10%~85%
Operating Conditions:
Environmental Conditions:
Temperature: 15
°C~35°C
Relative Humidity: 30%~60%
Atmospheric Pressure: 70kPa~106kPa
Power Supply Conditions:
Voltages: 380/400/420/440/460/480V,3~
Frequency: 50/60Hz
Input Current: 195A (Momentary), 45A (Continuous)
Power Output:
80kW/65kW/50kW Configurable
Max. Output:
When Generator is configured as 80kW: 80kW(1000mA@80kV)
When Generator is configured as 65kW: 65kW(500mA@130kV)
When Generator is configured as 50kW: 50kW(400mA@125kV)
Nominal Output:
When Generator is configured as 80kW: 80kW(800mA@100kV, 0.1s)
When Generator is configured as 65kW: 65kW(630mA@100kV, 0.1s)
When Generator is configured as 50kW: 50kW(500mA@100kV, 0.1s)
X-Ray Tube Voltage:
Digital Adjustable
Adjustment Range: 40kV~150kV