Appendix B: Specifications
5495975-1EN Rev.9 B-2
© 2013-2017 General Electric Company. All rights reserved.
Minimum Increment: 1kV
Allowable Deviation: 10%
X-Ray Tube Current:
Digital Adjustable in Multi-step pattern
Adjustment Range:
When configured as 80kW: 10mA~1000mA (totally 21 steps)
When configured as 65kW: 10mA~800mA (totally 20 steps)
When configured as 50kW: 10mA~630mA (totally 19 steps)
Allowable Deviation: 20%
Loading Time:
Exposure Time Range: 2.0ms~2s
Allowable Deviation: (10%+1ms)
X-Ray Tube Current Time Product:
Digital Adjustable in Multi-step pattern
Adjustment Range: 0.25mAs~630mAs
Allowable Deviation:
± (10%+0.2mAs)
AEC (Automatic Exposure Control) Mode:
KVp Range: 40~150kV
Max. AEC Backup Parameter: 512mAs and/or 2S
Nominal AEC Exposure Time: 5mS
Accuracy of Air Kerma Area Product Indicator:
±30%
Length Indicator:
SID:
Deviation between test data and indicated data is within
±5% (Not applicable to Simple Wall-
stand NBS2100)
Angulation:
Wallstand Detector Tray: Deviation between test data and indicated data is within ±2° (Not
applicable to Simple Wallstand NBS2100)
X-Ray Tube: Deviation between test data and indicated data is within ±2°
.
Digital Table (5176260 /GCTBL-C1):
Height from Tabletop to floor: 575mm~820mm (Allowable Deviation: ±10mm)
Table Detector Travel Range: 700mm (Allowable Deviation: ±10mm)
Minimum Height of Tabletop: 575mm (Allowable Deviation: ±10mm)
Tabletop Travel Range:
Horizontal Lateral: 232mm ±10mm
Horizontal Longitudinal: 770mm ±10mm
Digital Table (GCTBL-C2):
Height from Tabletop to floor: 500mm~850mm (Allowable Deviation: ±10mm)
Table Detector Travel Range: 850mm (Allowable Deviation: ±10mm)