Digital Wallstand:
The braking force of the linear movement 100N
The braking force of the rotation movement 100N
Manual Wallstand:
The braking force of the linear movement 100N
OTS:
The braking force of the lateral movement 100N
The braking force of the longitudinal movement 100N
Load Bearing:
Digital Table (5176260/GCTBL-C1): Maximum Weight Capacity is 220kg
Digital Table (GCTBL-C2): Maximum Weight Capacity is 320kg
Standard Table: Maximum Weight Capacity is 250kg
Mobile Table (5136793): Maximum Weight Capacity is 220kg
Digital Wallstand Detector (Non-Patient Weight Supporting Device): When at horizontal position,
supporting weight 30kg
Digital Detector: Maximum Weight Capacity is 160kg.
Noise:
Noise 70dB(A) in non-loading working status (not including the non-continuous or non-periodic
noise within 3 seconds.)
Generator/Tube Assembly Parameter under the intermittent operating mode:
Nominal X-Ray Tube Voltage: 150kV (with Current 500mA available: 500mA@150kV)
Maximum X-Ray Tube Current: 1000mA (with Voltage 80kV available: 80kV@1000mA)
Minimum Current Time Product: 0.5mAs
Nominal Minimum Exposure Time: 2ms