Appendix B: Specifications
5495975-1EN Rev.9 B-4
© 2013-2017 General Electric Company. All rights reserved.
– Digital Wallstand:
The braking force of the linear movement ≥100N
The braking force of the rotation movement ≥100N
– Manual Wallstand:
The braking force of the linear movement ≥100N
– OTS:
The braking force of the lateral movement ≥100N
The braking force of the longitudinal movement ≥100N
Load Bearing:
– Digital Table (5176260/GCTBL-C1): Maximum Weight Capacity is 220kg
– Digital Table (GCTBL-C2): Maximum Weight Capacity is 320kg
– Standard Table: Maximum Weight Capacity is 250kg
– Mobile Table (5136793): Maximum Weight Capacity is 220kg
– Digital Wallstand Detector (Non-Patient Weight Supporting Device): When at horizontal position,
supporting weight ≥ 30kg
– Digital Detector: Maximum Weight Capacity is 160kg.
Noise:
– Noise≤ 70dB(A) in non-loading working status (not including the non-continuous or non-periodic
noise within 3 seconds.)
Generator/Tube Assembly Parameter under the intermittent operating mode:
– Nominal X-Ray Tube Voltage: 150kV (with Current 500mA available: 500mA@150kV)
– Maximum X-Ray Tube Current: 1000mA (with Voltage 80kV available: 80kV@1000mA)
– Minimum Current Time Product: 0.5mAs
– Nominal Minimum Exposure Time: 2ms