Illustration 2: Normal Waveform
Illustration 3: Noisy ECG Trace (extra trigger at #4)
Illustration 4: Elevated T-wave (causing double trigger)
Illustration 5: ECG Trace (premature beat arrhythmia at trigger #3)
12.
Complete these activities to improve the ECG trace, if needed.
Check the cable connections.
If the external ECG trigger device is selected, position it away from the table and gantry
to minimize sensitivity to noise.
Consider an alternative electrode placement: remove electrodes, prep the patient again,
apply fresh electrodes, and repeat the impedance check.
Check default electrode selection on the active ECG trigger device.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
336 2 Cardiac Workflow