Table 3: Scan Set Up Operator Messages
Number English Operator Message Translated Operator Message
1
Patient orientation/position has changed from the previous
series.
Please verify the orientation/position of this series is as in‐
tended or change the orientation/position, if needed.
2
The table landmark has been changed. This changes the
location of all scans you have prescribed. Double check all
scans locations before you start scanning.
3
Dynamic Transition in SmartPrep is ON. System will auto‐
matically transition into Scan Phase when enhancement
level for Transition ROI reaches the HU value set for
threshold. Please confirm that the Transition ROI is set ap‐
propriately.
4
Image generation failed, scanning is paused: Use the
elapsed time to determine when to select Begin Scan Now.
5
The baseline location has changed.
You may proceed, but ROIs and dynamic transition will be
disabled. If you wish to use ROIs, please select [Retake
Baseline] to acquire a new baseline image.
6
The baseline location has changed. Dynamic Transition is
disabled.
Use the Monitor images to determine the transition or se‐
lect [Revisit Settings] to start a new SmartPrep series.
7
Injector is not armed. Please arm the injector or select Dis‐
able the Injector from the Contrast collection.
8
The injector is not communicating or is not in the right state.
Please cycle power on the injector or disable the Injector in
the Contrast collection.
If the problem persists, contact your service representative.
9
There is a mismatch with the Injector settings. Please
check the injection settings in the Contrast collection.
10
Please select Revisit Settings and check Injector settings or
disable the Injector from the Contrast collection.
11
Injector is not ready. Please cycle power on the injector or
select Injector Off to disable the Injector from the Contrast
collection.
12
Internal communication error.
Please perform a Scan Hardware Reset from the Service
Tools menu.
If problem persists, contact your service representative.
13
System detected internal communication error. Please se‐
lect [Done Scanning] or [Close Exam] and perform a Scan
Hardware Reset from the Service Tools menu.
If problem persists, contact your service representative.
14
ODM has been disabled. Verify Noise Index, Min mA/Max
mA, resultant CTDIvol and DLP meet the clinical goal of the
acquisition and follow the radiation exposure guidelines for
your department.
15
Projected noise index is less than the prescribed noise in‐
dex. Review and reduce minimum mA, if appropriate.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
670 1 Operator Messages