Troubleshooting
84 SEER™ 1000 2067634-077 Revision B
Trouble Indicators
Cause
Recommendation
The Event Button LED is
a solid yellow.
Previous recording or patient
entry has not been deleted.
To start a new recording, de-
lete the recording and/or pa-
tient demographic data using
either the Holter analysis sys-
tem software or the SEER 1000
app.
The Event Button LED
stops flashing green
during specified record-
ing duration.
The recording has stopped
before the specified record-
ing duration has been
reached.
The recording may have been
started with a low battery.
Download or delete the record-
ing and replace the battery to
start a new recording.
Check the unit for physical
damage. If the unit is dam-
aged, contact service.
The Event Button LED
and the Hookup Guid-
ance LED flash green
and yellow alternating
three times every sec-
ond.
Recorder cannot be started.
Self-test failed.
See section How to Notify Ser-
vice if the Recorder is Defec-
tiveon page 91.