GE Analytical Instruments ©2010 126 of 226 DLM 74001-04 Rev. A
Chapter 6: Calibration and System Protocols
11. Press the Exit.
12. If you have a Super iOS system, and selected the After or the Before and After option, the Analyzer will
prompt you to remove the standards, and to insert the rinse cartridge or vials into the Super iOS system
to continue the Rinse activity. Take out the rinse cartridge or vials when completed.
13. If you have a standard iOS System, remove the standards and slide the iOS door closed.
Figure 29: The Accuracy, Precision, and Verification Summary Screen, Part 1
Figure 30: The Accuracy, Precision, and Verification Summary Screen, Part 2
Relative Standard Deviation (RSD) =
Standard Deviation
Measured TOC Concentration
--------------------------------------------------------------------------
100