GE Analytical Instruments ©2009 163 of 236 DLM 90488-01 Rev. A
Calibration pass/fail based upon the following criteria:
•The R
2
is 0.990.
The % Diff value for each IC standard is ±10% or less.
Calibration Results:
Passed Failed
Calibration Action:
Applied Canceled
% Diff =
Measured Concentration - Expected Standard Concentration
Ex
p
ected Standard Concentration
-----------------------------------------------------------------------------------------------------------------------------------------------------------------
100%×