GE Analytical Instruments ©2009 164 of 236 DLM 90488-01 Rev. A
Verification Worksheet
Company Name___________________________ Date ________________________
Analyst Name ____________________________ Firmware Version _____________
Analyzer Serial Number ____________________
Concentration of Standard: _______
(Line 1)
Reagent Flush Performed (Y/N): ______ (Line 2)
Enter data for the TOC Verification:
Expected Measured RSD % Diff
Blank N/A
________ ________ ________
________ ________ ________ ________ ________
Enter data for the IC Verification:
Expected Measured RSD % Diff
________ ________ ________ ________ ________
Pass criteria:
• RSD for each vial is 3% or less.
• TOC % Diff is ±7% or less for a 500 ppb or 1 ppm TOC standard.
•TOC % Diff is ±5% or less for a 2, 5, 10, 25, or 50 ppm TOC standard.
• IC % Diff is ±7 or less% for a 500 ppb or 1 ppm IC standard.
•IC % Diff is ±5% or less for a 2, 5, 10, 25, or 50 ppm IC standard.
Verification Results:
Passed Failed
% Diff =
Measured Concentration - Expected Standard Concentration
Ex
ected Standard Concentration
-----------------------------------------------------------------------------------------------------------------------------------------------------------------
100%×