GE Analytical Instruments ©2009 166 of 236 DLM 90488-01 Rev. A
System Suitability Worksheet
Company Name___________________________ Date ________________________
Analyst Name ____________________________ Firmware Version _____________
Analyzer Serial Number ____________________
Average TOC of Reagent Water (Rw) __________________
(Line 1)
Average TOC of 500 ppb Sucrose Standard (Rs) __________________ (Line 2)
Average TOC of 500 ppb Benzoquinone Standard (Rss) __________________ (Line 3)
Response Efficiency __________________ (Line 4)
Response Limit __________________ (Line 5)
Response Limit = Rs - Rw
Acceptance criteria for USP System Suitability is response efficiency between 85% and 115%
Pass Fail
Indicate whether the response efficiency results were applied or canceled:
Apply Cancel
Response Efficiency =
(Rss - Rw)
(Rs-Rw)
----------------------------
100×