NOTE: 1. Must locate center of the magnet to keep minimum service area described in
Illustration 2-2.
2.
Minimum dimensions are for service only. Room size may grow due to local
constraints including magnetic field containment
3.
Minimum Magnet Room dimensions do not contain 5 gauss line to room. If fringe
field containment is needed, see Chapter 3, Magnetic Shielded Room
Requirement .
4. Magnet Room minimum ceiling height is 98.5 in. (2500mm) when M7000GM 2.5m
Ceiling Kit-Passive is installed. If suspended ceiling height is between 98.5 in.
(2500 mm) and 105 in. (2667 mm), 2.5M Low Ceiling Kit-Passive, M7000GM is
required. Contact the PMI for preparation.
SIGNA Voyager Pre-Installation
Direction 5680008–1EN, Revision 2
24 2 MR Suite Minimum Room Size Requirement